Parameter | 36-1: Phycocyanin (PC) 36-2: Phycoerythrin (PE) |
Measurement range | 36-1: 0 – 4,500 ppb (μg/L), 0~150,000 cells/ml 36-2: 0 – 4,500 ppb (μg/L), 0~450,000 cells/ml |
Accuracy | ± 3% FS |
Resolution | 0.01 ppb |
Repeatability | ± 1% FS |
Light source | 36-1: LED 590 nm 36-2: LED 525 nm |
Operation pressure | Max. 5 Kg f/cm2 |
Operating Temperature | 0 – 85 °C |
Process flow rate | Min. 0.1 m/sec |
Response time | 3 sec |
Protection | IP 68 |
Temperature sensor | Pt 100 |
Temperature measurement range | 0 – 85°C |
Temperature accuracy | ± 0.5 °C ; ± 0.1 °C (Optional) |
Housing | SS 316 L ; Titanium |
Dimension | 35 Χ 215 mm |
Weight | Analyzer: approx. 500 g; Cable: 80 g/m |