Conductivity Analyzer Cakrawala Bima Instrument


  • Water submersible enclosurea and solution without cabinet
  • Temperature compensation
  • On line real time measurement 1 second
  • On board memory allowing users to easily configure, calibrate sensor at lab, and distribute to various fields sites
  • 05 W ultra low power consumption and ideal for out door applications
  • Lightening and surge protection for power
  • High reliability and drift free
  • Digital communication, minimize cabling, and engineering cost
  • Data monitoring and software configuration calibration
  • Standard modbus RTU protocol, direct connected with PLC , HMI , and eliminate I/O module cost

Measurement range 0 – 5,000 μS/cm;

0 – 1,000/10,000/50,000 μS/cm (auto range);

0 – 10,000/100,000/500,000 μS/cm (auto range);

0 – 2,000/20,000/100,000 μS/cm(auto range);

0 – 20/200/1,000 μS/cm (auto range);

0 – 100/1,000/5,000 μS/cm (auto range);

0 – 20,000/200,000/1000,000 μS/cm (auto range): (Optional)

Accuracy ± 0.5% FS
Resolution 0.1 μS/cm
Repeatability ± 0.5% FS
Operation pressure Max. 10 Kg f/cm2
Response time 1 sec
Protection IP 68
Temperature sensor PT 100
Temperature measurement range 0 – 85 °C; 0 – 100 °C (Optional)
Temperature accuracy ± 0.5 °C; ± 0.1 °C (Optional)
Process connection M 28 x 1.5
Housing SS 316 L; POM; Titanium
Dimension 25 Χ 210 mm
Weight