Incubator Laboratorium – Type INC-100 (Local)
Juni 21, 2019
Optimus Yellow - Sound Level Meter
Optimus Red – Sound Level Meter
Juli 3, 2019

doseBadge5 – Noise Dosimeter

NEW doseBadge5 Noise Dosimeter

NEW doseBadge5 Noise Dosimeter

 

DosBadge5 adalah dosimeter kebisingan nirkabel kinerja tinggi yang memungkinkan Anda untuk membuat pengukuran paparan kebisingan yang sesuai dengan pekerjaan, menerapkan langkah-langkah kontrol yang efektif dan memastikan bahwa pekerja tidak terpapar pada tingkat kebisingan yang akan membuat pendengaran mereka dalam risiko.

 

Standards
  • IEC 61252:1993 +AMD1:2000 Personal Sound Exposure Meters
  • IEC 61252 Ed 1.1 (2002-03) Personal Sound Exposure Meters
  • ANSI S1.25:1991 (R2017) Personal Noise Dosimeters
  • IEC 61260-1:2014 Class 2 (Octave Bands from 63Hz to 8kHz)
  • ANSI S1.11-2014 Class 2 (Octave Bands from 63Hz to 8kHz)
Measurement Range
  • RMS Range: 60 dB – 140 dB (A)
  • Peak Range: 80 dB – 143 dB (C)
  • Octave Bands: 70 dB – 140 dB (A)
Measurement Functions
  • 4 Simultaneous Independent Integrator Channels
  • 2 Simultaneous Independent Peak Channels
  • 1:1 Octave Bands (63Hz to 8kHz)
Overall Measurement Parameters
  • Measurement duration, Start time & date, Instrument serial number & name
  • Calibration information (field & factory calibration), Overload & Tamper sensor detection
Overall Integrator Parameters
  • Average Integrated Sound Level (Leq/LAVG), Time Weighted Average (LEP,d/LEX,8/TWA)
  • Sound Exposure & Estimated Sound Exposure, % Dose & Estimated % Dose, ULT Duration
  • SPL Max Level & Time, SPL Min Level
Overall Peak Channel Parameters
  • Overall LPeak level for each channel (Up to 2 channels)
Overall Octave Band Parameters
  • Overall unweighted Leq for each octave band (63Hz to 8kHz)
Communication
  • Bluetooth® Wireless communication to the doseBadge5 dBControl
  • Bluetooth® Wireless communication to the dBLink App (Android & iOS)
  • USB download to NoiseTools via the doseBadge5 Dock
Integrator Channel Options
  • Channel name – Preset or user defined
  • Exchange rate – 3, 4 or 5 dB
  • Criterion Level – 80 dB to 100 dB in 1 dB steps
  • Criterion Time – 1 hr to 24 hrs in 1hr steps
  • Threshold Level – None, 70 dB to 100 dB n 1dB steps
  • Time Weighting – Fast, Slow or None
  • Frequency Weighting – A, C or Z
  • ULT Level – 70 dB to 140 dB in 1 dB steps
  • ULT Time Weighting – None, Fast or Slow
  • ULT Freq. Weighting – A,C, or Z
Peak Channel Options
  • Frequency weighting – A,C or Z (2 channels)
Calibration
  • Automatic detection of external acoustic calibration
  • User configurable calibration level (typically 114dB or 94dB)
Data Logging
  • 1 second – 1 minute time history data (user selectable)
Memory
  • Up to 80 hours of Time History Data for each channel including 1:1 Octave Bands
  • Up to 40 individual measurements
  • Maximum duration of any single measurement: 24 hours
Dimensions
  • 66 x 43 x 53 mm (excluding clips)
  • 2.53” x 1.69” x 2.01” (excluding clips)
  • 85 g/2.9 oz

  • ATEX, EEx, IECEx & FM Certification for hazardous atmospheres.

 

Harga: (hubungi kami)

Bagikan:

Comments are closed.